Description

In vivo dosimetry was performed for the head and neck carcinoma patients during the treatment of a large photon field using MOSFETs. This study followed the protocols recommended by the European Society for Therapeutic Radiology and Oncology. A total of 32 portals belonging to 12 patients were under investigation. Results showed that the deviation between in vivo midline doses and planned target doses was partly due to the manual dose calculations in the treatment planning which used the patient geometric thickness rather than the radiological thickness. Other factors responsible for this deviation included the difficult positioning of MOSFETs on the face mask, the asymmetric positioning of MOSFETs on the left and right sides of the mask, and the asymmetric tissue inhomogeneities with respect to the body midline. To reduce the deviation contributed from these factors, in vivo midline doses were calculated by averaging the results for each bilaterally opposed portals and compared with corresponding planned target doses. This comparison showed that MOSFET dosimeters are suitable for in vivo dosimetry of the present study

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radmeas.2007.11.050

Additional details

Identifiers

DOI
10.1016/j.radmeas.2007.11.050;
PII
S1350-4487(07)00523-9;

Publishing Information

Journal Title
Radiation Measurements
Journal Volume
43
Journal Issue
2-6
Journal Page Range
p. 870-874
ISSN
1350-4487
CODEN
RMEAEP

Conference

Title
15. solid state dosimetry conference
Acronym
SSD15
Dates
8-13 Jul 2007
Place
Delft (Netherlands)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
40008533
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CARCINOMAS; DOSEMETERS; DOSIMETRY; IN VIVO; MOSFET; NECK; PATIENTS; PHOTONS; QUALITY ASSURANCE; RADIATION DOSES; RESPIRATORS; THICKNESS; VERIFICATION
Descriptors DEC
BODY; BOSONS; DIMENSIONS; DISEASES; DOSES; ELEMENTARY PARTICLES; FIELD EFFECT TRANSISTORS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; MOS TRANSISTORS; NEOPLASMS; SEMICONDUCTOR DEVICES; TRANSISTORS

Optional Information

Copyright
Copyright (c) 2007 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.