In vivo dose verification for photon treatments of head and neck carcinomas using MOSFET dosimeters
- 1. Department of Biomedical Engineering and Environmental Sciences, National Tsing Hua University, Hsinchu, Taiwan (China)
- 2. Department of Radiation Oncology, Kaohsiung Veterans General Hospital, Kaohsiung, Taiwan (China)
- 3. Department of Medical Imaging and Radiological Sciences, Chang Gung University, Kwei-Shan, Taiwan (China)
Description
In vivo dosimetry was performed for the head and neck carcinoma patients during the treatment of a large photon field using MOSFETs. This study followed the protocols recommended by the European Society for Therapeutic Radiology and Oncology. A total of 32 portals belonging to 12 patients were under investigation. Results showed that the deviation between in vivo midline doses and planned target doses was partly due to the manual dose calculations in the treatment planning which used the patient geometric thickness rather than the radiological thickness. Other factors responsible for this deviation included the difficult positioning of MOSFETs on the face mask, the asymmetric positioning of MOSFETs on the left and right sides of the mask, and the asymmetric tissue inhomogeneities with respect to the body midline. To reduce the deviation contributed from these factors, in vivo midline doses were calculated by averaging the results for each bilaterally opposed portals and compared with corresponding planned target doses. This comparison showed that MOSFET dosimeters are suitable for in vivo dosimetry of the present study
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radmeas.2007.11.050Additional details
Identifiers
- DOI
- 10.1016/j.radmeas.2007.11.050;
- PII
- S1350-4487(07)00523-9;
Publishing Information
- Journal Title
- Radiation Measurements
- Journal Volume
- 43
- Journal Issue
- 2-6
- Journal Page Range
- p. 870-874
- ISSN
- 1350-4487
- CODEN
- RMEAEP
Conference
- Title
- 15. solid state dosimetry conference
- Acronym
- SSD15
- Dates
- 8-13 Jul 2007
- Place
- Delft (Netherlands)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 40008533
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CARCINOMAS; DOSEMETERS; DOSIMETRY; IN VIVO; MOSFET; NECK; PATIENTS; PHOTONS; QUALITY ASSURANCE; RADIATION DOSES; RESPIRATORS; THICKNESS; VERIFICATION
- Descriptors DEC
- BODY; BOSONS; DIMENSIONS; DISEASES; DOSES; ELEMENTARY PARTICLES; FIELD EFFECT TRANSISTORS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; MOS TRANSISTORS; NEOPLASMS; SEMICONDUCTOR DEVICES; TRANSISTORS
Optional Information
- Copyright
- Copyright (c) 2007 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.