The Use of TLD and Gafchromic Film to Assure Submillimeter Accuracy for Image-Guided Radiosurgery
- 1. Theresa Po Cyberknife Center, Hong Kong Adventist Hospital, Hong Kong (China)
- 2. Department of Radiation Oncology, Stanford University Medical Center, Stanford, CA (United States)
- 3. Department of Neurosurgery, Stanford University Medical Center, Stanford, CA (United States)
- 4. Accuray, Inc., Sunnyvale, CA (United States)
Description
The Cyberknife is an image-guided radiosurgical system. It uses a compact X-band 6-MV linear accelerator mounted on a robotic arm to deliver radiosurgical doses. While routine quality assurance (QA) is essential for any radiosurgery system, QA plays an even more vital role for the Cyberknife system, due to the complexity of the system and the wide range of applications. This paper presents a technique for performing quality assurance using thermoluminescence detectors (TLDs) and Gafchromic films that is intended to be specific for the Cyberknife. However, with minor modification, the proposed method can also be used for QA of other radiosurgery systems. Our initial QA procedure for the CyberKnife utilized a 30 x 30 x 11-cm solid water phantom containing a planar array of slots for 1x 1 x 1-mm TLDs on a 2-mm grid. With the objective of significantly simplifying CyberKnife QA, a new procedure for verification was developed, which uses much fewer TLDs than the prior solid water phantom technique. This new method requires only that the system target dose to the center of a cluster of 7 TLDs. In a prior study with Gafchromic films, conducted at 3 different Cyberknife facilities, the mean clinically relevant error was demonstrated to be 0.7 mm. A similar Gafchromic film analysis replicated these error measurements as part of the present investigation. It cannot be emphasized enough the importance of implementing routine QA to verify the accuracy of any radiosurgery system. Our quality assurance procedure tests the treatment planning system, as well as the entire treatment delivery including the image targeting system and the robot system. Either TLDs or Gafchromic films may be used for QA test of a radiosurgery system. Using both methods for measurement has the advantage independently verifying the accuracy of the system. This approach, which is routinely in used at our institution, has repeatedly confirmed the submillimeter targeting accuracy of our Cyberknife
Availability note (English)
Available from http://dx.doi.org/10.1016/j.meddos.2007.04.009Additional details
Identifiers
- DOI
- 10.1016/j.meddos.2007.04.009;
- PII
- S0958-3947(07)00106-9;
Publishing Information
- Journal Title
- Medical Dosimetry
- Journal Volume
- 33
- Journal Issue
- 1
- Journal Page Range
- p. 36-41
- ISSN
- 0958-3947
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 39064674
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ACCURACY; IMAGES; LINEAR ACCELERATORS; PHANTOMS; QUALITY ASSURANCE; RADIATION DOSES; RADIOTHERAPY; SURGERY; THERMOLUMINESCENT DOSEMETERS; THERMOLUMINESCENT DOSIMETRY
- Descriptors DEC
- ACCELERATORS; DOSEMETERS; DOSES; DOSIMETRY; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIOLOGY; STRUCTURAL MODELS; THERAPY
Optional Information
- Copyright
- Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.