Published 1995 | Version v1
Journal article

Autoblocking dose-limiting normal structures within a radiation treatment field: 3-D computer optimization of 'unconventional' field arrangements

Description

Purpose/Objective: To demonstrate that one can obtain a homogeneous dose distribution within a specified gross tumor volume (GTV) while severely limiting the dose to a structure surrounded by that tumor volume. We present three clinical examples below. Materials and Methods: Using planning CT scans from previously treated patients, we designed variety of radiation treatment plans in which the dose-critical normal structure was blocked, even if it meant blocking some of the tumor. To deal with the resulting dose inhomogeneities within the tumor, we introduced 3D compensation. Examples presented here include (1) blocking the spinal cord segment while treating an entire vertebral body, (2) blocking both kidneys while treating the entire peritoneal cavity, and (3) blocking one parotid gland while treating the oropharynx in its entirety along with regional nodes. A series of multiple planar and non-coplanar beam templates with automatic anatomic blocking and field shaping were designed for each scenario. Three-dimensional compensators were designed that gave the most homogeneous dose-distribution for the GTV. For each beam, rays were cast from the beam source through a 2D compensator grid and out through the tumor. The average tumor dose along each ray was then used to adjust the compensator thickness over successive iterations to achieve a uniform average dose. DVH calculations for the GTV, normal structures, and the 'auto-blocked' structure were made and used for inter-plan comparisons. Results: These optimized treatment plans successfully decreased dose to the dose-limiting structure while at the same time preserving or even improving the dose distribution to the tumor volume as compared to traditional treatment plans. Conclusion: The use of 3D compensation allows one to obtain dose distributions that are, theoretically, at least, far superior to those in common clinical use. Sensible beam templates, auto-blocking, auto-field shaping, and 3D compensators form a powerful 3D optimization package

Additional details

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
32
Journal Issue
971
Journal Page Range
p. 300
ISSN
0360-3016
CODEN
IOBPD3

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34059333
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
NEOPLASMS; OPTIMIZATION; PLANNING; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; VOLUME
Descriptors DEC
DISEASES; DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY

Optional Information

Copyright
Copyright (c) 1995 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.