Published May 1991 | Version v1
Journal article

A new applicator, positionable to the center of tracheobronchial lumen for HDR-IR-192-afterloading of tracheobronchial tumors

  • 1. Univ. of Heidelberg, (F.R.Germany)

Description

A new bronchial applicator for afterloading irradiation is introduced which can be positioned to the center of the tracheobronchial lumen. The central position in the lumen leads to a clear improvement of dose distribution. The applicator is built on the principle of a coaxial tube. Parts of the outer cover can be expanded to baskets and effect a distance of the radiation source from the bronchial mucosa or tumor surface, and at the same time, expend a relief of extreme contact doses. No obstruction of the respiratory system through the positioning device will be caused. The positionable bronchial applicator seems to be suitable for reducing complications caused through high contact doses and irregular dose distributions and may be able to improve the results of endoluminal radiotherapy

Additional details

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
20
Journal Issue
5
Series
Int. J. Radiat. Oncol., Biol. Phys.
Journal Page Range
1061-1066
ISSN
0360-3016
CODEN
IOBPD

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
22077145
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
AFTERLOADING; BRONCHI; IRIDIUM ISOTOPES; NEOPLASMS; POSITIONING; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; TRACHEA
Descriptors DEC
DISEASES; MEDICINE; RESPIRATORY SYSTEM; THERAPY