Gamma-ray activity determination in large volume samples with a Ge-Li detector
Description
A method for the determination of the γ-ray activity of large volume sources is described and discussed. The efficiency of the detector is expressed as the product of an intrinsic, a geometrical, and a self-absorption factor. The first and the second terms are determined experimentally while the last is evaluated theoretically. The method has been applied to the determination of the γ specific activity of natural potassium and natural lanthanum. For potassium a value of 3.21 +- 0.01 γ/s g of K was obtained, while for 138La the partial half-lives for the photons of 789.1 and 1435.8 keV turned out to be 3.68 +- 0.14 x 1011 y and 1.99 +- 0.03 x 1011 y, respectively
Additional details
Identifiers
- DOI
- 10.1021/ac50016a024;
Publishing Information
- Journal Title
- Analytical Chemistry
- Journal Volume
- 49
- Journal Issue
- 8
- Series
- Anal. Chem.
- Journal Page Range
- 1156-1159
- ISSN
- 0003-2700
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 8345996
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY; S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- EFFICIENCY; GAMMA DETECTION; GAMMA SPECTRA; LANTHANUM; LI-DRIFTED GE DETECTORS; POTASSIUM; RADIOMETRIC ANALYSIS
- Descriptors DEC
- ALKALI METALS; CHEMICAL ANALYSIS; ELEMENTS; GE SEMICONDUCTOR DETECTORS; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; METALS; QUANTITATIVE CHEMICAL ANALYSIS; RADIATION DETECTION; RADIATION DETECTORS; RARE EARTHS; SEMICONDUCTOR DETECTORS; SPECTRA
Optional Information
- Notes
- Updated automatically by Metadata and Full-Text Enrichment Agent