Published 1988
| Version v1
Report
A neutron dosimetry intercomparison study with tissue equivalent ionization chambers: National Accelerator Centre and Universite Catholique de Louvain, Belgium
Description
The results of a neutron dosimetry intercomparison study undertaken at the Universite Catholique de Louvain, Belgium are reported. The results show excellent agreement between the different tissue equivalent ionization chambers used. The photon calibration factors measured differ from those previously determined by a maximum of 0.14%. For the measurement of total dose in a p(65 MeV)/Be(35 MeV) neutron therapy beam the maximum deviation from the measured mean is 0.48% for all conditions. For the NAC chamber used the photon calibration differed by 0.11% from the measurement made at NAC, while the average deviation from the mean for the neutron beam measurements was 0.10%
Availability note (English)
Available from the CSIR, National Accelerator Centre, P.O. Box 72, Faure, 7131, South Africa.Additional details
Publishing Information
- ISBN
- 0 7988 4183 4
- Imprint Pagination
- 30 p.
- Report number
- NAC--88-02
INIS
- Country of Publication
- South Africa
- Country of Input or Organization
- South Africa
- INIS RN
- 20049114
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Numerical Data, Non-conventional Literature
- Descriptors DEI
- BRAGG GRAY CHAMBERS; CALIBRATION; COMPARATIVE EVALUATIONS; EXPERIMENTAL DATA; NEUTRON BEAMS; NEUTRON DOSIMETRY; NEUTRON THERAPY; PHOTONS; RADIATION DOSES
- Descriptors DEC
- BEAMS; BOSONS; DATA; DOSEMETERS; DOSIMETRY; ELEMENTARY PARTICLES; EVALUATION; INFORMATION; IONIZATION CHAMBERS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; MEDICINE; NUCLEON BEAMS; NUMERICAL DATA; PARTICLE BEAMS; RADIATION DETECTORS; RADIOTHERAPY; THERAPY