Published 1988 | Version v1
Report

A neutron dosimetry intercomparison study with tissue equivalent ionization chambers: National Accelerator Centre and Universite Catholique de Louvain, Belgium

Description

The results of a neutron dosimetry intercomparison study undertaken at the Universite Catholique de Louvain, Belgium are reported. The results show excellent agreement between the different tissue equivalent ionization chambers used. The photon calibration factors measured differ from those previously determined by a maximum of 0.14%. For the measurement of total dose in a p(65 MeV)/Be(35 MeV) neutron therapy beam the maximum deviation from the measured mean is 0.48% for all conditions. For the NAC chamber used the photon calibration differed by 0.11% from the measurement made at NAC, while the average deviation from the mean for the neutron beam measurements was 0.10%

Availability note (English)

Available from the CSIR, National Accelerator Centre, P.O. Box 72, Faure, 7131, South Africa.

Additional details

Publishing Information

ISBN
0 7988 4183 4
Imprint Pagination
30 p.
Report number
NAC--88-02

INIS

Country of Publication
South Africa
Country of Input or Organization
South Africa
INIS RN
20049114
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Numerical Data, Non-conventional Literature
Descriptors DEI
BRAGG GRAY CHAMBERS; CALIBRATION; COMPARATIVE EVALUATIONS; EXPERIMENTAL DATA; NEUTRON BEAMS; NEUTRON DOSIMETRY; NEUTRON THERAPY; PHOTONS; RADIATION DOSES
Descriptors DEC
BEAMS; BOSONS; DATA; DOSEMETERS; DOSIMETRY; ELEMENTARY PARTICLES; EVALUATION; INFORMATION; IONIZATION CHAMBERS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; MEDICINE; NUCLEON BEAMS; NUMERICAL DATA; PARTICLE BEAMS; RADIATION DETECTORS; RADIOTHERAPY; THERAPY