Published 1990
| Version v1
Miscellaneous
Bone mineral content evaluation system: calibration
- 1. Universidade Federal, Rio de Janeiro, RJ (Brazil). Coordenacao dos Programas de Pos-graduacao de Engenharia
Description
We present in this paper a device developed in our laboratory for evaluating the bone mineral content in vivo by using monoenergetic gamma rays. The method employed is based upon the attenuation of a low energy radiation beam, being characterized by measurements of the intensity beam, performed at equally spaced intervals in a specified forearm cross section area. The precision obtained with this procedure (in terms of bone mineral content) is about five percent. The total measurement time is 15 minutes and the radiation dose the patient receive is considerably low and restricted to a narrow band of the forearm. (author)
Additional details
Additional titles
- Original title (Portuguese)
- Sistema de medida do conteudo de mineral osseo: calibracao
Publishing Information
- Imprint Title
- Proceedings of the 3. General Congress on Nuclear Energy. v. 8
- Imprint Pagination
- 123 p.
- Journal Page Range
- p. 97-105.
Conference
- Title
- 3. General Congress on Nuclear Energy.
- Dates
- 22-27 Jul 1991.
- Place
- Rio de Janeiro, RJ (Brazil).
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 23087608
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- ARMS; CALIBRATION; GAMMA RADIATION; IN VIVO; MINERALS; SKELETON
- Descriptors DEC
- BODY; BODY AREAS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; LIMBS; ORGANS; RADIATIONS
Optional Information
- Notes
- Imprint:Anais do 3. Congresso Geral de Energia Nuclear. v. 8.