Published 2009 | Version v1
Miscellaneous Open

Performance of X-ray equipment used for conventional radiographic examinations of children

  • 1. Centro de Desenvolvimento da Tecnologia Nuclear (CDTN-CNEN/MG), Belo Horizonte, MG (Brazil)
  • 2. Universidade Federal de Pernambuco (UFPE), Recife, PE (Brazil). Dept. de Energia Nuclear

Description

The performance of X-ray equipment that are used for conventional X-ray examinations of children in four hospitals of Belo Horizonte, Brazil, was evaluated. The methodology proposed by the Brazilian authority for radiation protection in diagnostic radiology was adopted. The performance tests were carried out in six X-ray machines and they consisted of measurements and determination of the X-ray tube output, linearity and constancy of radiation output, accuracy and reproducibility of the exposure time and tube potential, half-value layer (HVL), light field / X-ray field alignment and accuracy of the X-ray field indicator. It was observed that only one hospital had the suitable equipment for X-ray examinations of children. Results showed that all six equipment of the four hospitals presented unsatisfactory performance in some quality control tests. Only for some combinations of the exposure parameters, generally employed in pediatric X-ray examinations, some equipment complied with the authority requirements. (author)

Availability note (English)

Available from INIS in electronic form

Files

40106330.pdf

Files (223.0 kB)

Name Size Download all
md5:0e2eb10de79ffdc6b519f54dec0975f1
223.0 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
[7 p.]
Report number
INIS-BR--6304

Conference

Title
International nuclear atlantic conference. Innovations in nuclear technology for a sustainable future; 16. Brazilian national meeting on reactor physics and thermal hydraulics; 9. Brazilian national meeting on nuclear applications; 1. Meeting on nuclear industry
Acronym
INAC 2009
Dates
27 Sep - 2 Oct 2009
Place
Rio de Janeiro, RJ (Brazil)

INIS

Country of Publication
Brazil
Country of Input or Organization
Brazil
INIS RN
40106330
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference, Numerical Data
Descriptors DEI
ACCURACY; BIOMEDICAL RADIOGRAPHY; CHILDREN; EXPERIMENTAL DATA; HOSPITALS; PERFORMANCE; QUALITY ASSURANCE; QUALITY CONTROL; RADIATION PROTECTION; X-RAY EQUIPMENT
Descriptors DEC
AGE GROUPS; ANIMALS; BUILDINGS; CONTROL; DATA; DIAGNOSTIC TECHNIQUES; EQUIPMENT; INFORMATION; MAMMALS; MAN; MEDICAL ESTABLISHMENTS; MEDICINE; NUCLEAR MEDICINE; NUMERICAL DATA; PRIMATES; RADIOLOGY; VERTEBRATES

Optional Information

Notes
4 refs., 6 figs., 3 tabs. Published only in CD-Rom. Code: e02_529_fullpaper.pdf