Published May 1986
| Version v1
Report
Restricted
Results from a test of a Cu-scintillator calorimeter module with photodiode readout
Description
A calorimeter module of 17 radiation lengths depth has been built. Wavelength shifter (WLS) bars coupled to rectangular silicon photodiodes (PD's) are use as readout. Considerations in the design of the WLS bars, with particular emphasis on optimising the efficiency for PD readout, are discussed. The energy resolution for electrons has been determined to be about 9%/√E between 2 and 50 GeV. The response to hadrons is presented and the prospects for the construction of a full-sized hadron calorimeter are discussed. (orig.)
Availability note (English)
MF available from INIS under the Report Number.
Files
Additional details
Publishing Information
- Imprint Pagination
- 28 p.
- Report number
- MPI-PAE/Exp.El.--163
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 17055532
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALORIMETERS; COPPER; EFFICIENCY; ELECTRON DETECTION; ENERGY DEPENDENCE; ENERGY RESOLUTION; GEV RANGE 01-10; GEV RANGE 10-100; OPTICAL SYSTEMS; OPTIMIZATION; PHOTODIODES; PION DETECTION; PLASTIC SCINTILLATION DETECTOR; READOUT SYSTEMS; RELATIVISTIC RANGE; RESPONSE FUNCTIONS; SCINTILLATOR-PHOTODIODE DETECT; SILICON DIODES; SPECTRAL SHIFT
- Descriptors DEC
- CHARGED PARTICLE DETECTION; DETECTION; ELEMENTS; ENERGY RANGE; FUNCTIONS; GEV RANGE; MEASURING INSTRUMENTS; METALS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION; SCINTILLATION COUNTERS; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; SOLID SCINTILLATION DETECTORS; TRANSITION ELEMENTS