Uncertainty evaluation of the kerma in the air, related to the active volume in the ionization chamber of concentric cylinders, by Monte Carlo simulation

  • 1. Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil). Lab. Nacional de Metrologia das Radiacoes Ionizantes (LNMRI)

Description

To implant the primary standard of the magnitude kerma in the air for X-ray between 10 - 50 keV, the National Metrology Laboratory of Ionizing Radiations (LNMRI) must evaluate all the uncertainties of measurement related with Victtoren chamber. So, it was evaluated the uncertainty of the kerma in the air consequent of the inaccuracy in the active volume of the chamber using the calculation of Monte Carlo as a tool through the Penelope software