Published June 1999 | Version v1
Journal article

Experimental research on the intensity distribution of KrF laser beam with large diameter and high energy

  • 1. China Inst. of Atomic Energy, Beijing (China)

Description

The array calorimeter detector was used to measure the uniformity of intensity distribution and output energy for large diameter (φ267 mm) and high energy (400 J) KrF laser beam. With a decrease of the diode voltage, the distribution becomes worse to have a saddle pattern. The optimum operation voltage of diode are of 580 to 600 kV

Additional details

Publishing Information

Journal Title
High Power Laser and Particle Beams
Journal Volume
11
Journal Issue
3
Journal Page Range
p. 257-261
ISSN
1001-4322

INIS

Country of Publication
China
Country of Input or Organization
China
INIS RN
31022861
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
BEAMS; CALORIMETERS; DIODE TUBES; DISTRIBUTION; ELECTRIC POTENTIAL; ENERGY; KRYPTON FLUORIDE LASERS
Descriptors DEC
ELECTRON TUBES; EXCIMER LASERS; GAS LASERS; LASERS; MEASURING INSTRUMENTS