Cerebral imaging using 68Ga DTPA and the U.C.S.F. multiwire proportional chamber positron camera
Creators
- 1. Univ. of California, San Francisco
Description
A multiwire proportional chamber positron camera consisting of four 48 x 48 cm2 detectors linked to a small digital computer has been designed, constructed, and characterized. Initial clinical application to brain imaging using 68Ga DTPA in 10 patients with brain tumors is described. Tomographic image reconstruction is accomplished by an algorithm determining the intersection of the annihilation photon paths in planes of interest. Final image processing utilizes uniformity correction, simple thresholding, and smoothing. The positron brain images were compared to conventional scintillation brain scans and x-ray computerized axial tomograms (CAT) in each case. The positron studies have shown significant mitigation of confusing superficial activity resulting from craniotomy in comparison to conventional brain scans. Central necrosis of lesions observed in the positron images, but not in the conventional scans, has been confirmed in CAT. Modifications of the camera are being implemented to improve image quality, and these changes combined with the tomography inherent in the positron scans are anticipated to result in images superior in information content to conventional brain scans
Additional details
Identifiers
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 23
- Journal Issue
- 1
- Series
- IEEE Trans. Nucl. Sci.
- Journal Page Range
- 523-527
- ISSN
- 0018-9499
Conference
- Title
- 22. nuclear science symposium and 7. nuclear power systems symposium.
- Dates
- 19 Nov 1975.
- Place
- San Francisco, CA, USA.
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 7261228
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BRAIN; GALLIUM 68; MULTIWIRE PROPORTIONAL CHAMBER; NEOPLASMS; PERFORMANCE; POSITRON CAMERAS; RADIOISOTOPE SCANNING
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-PLUS DECAY RADIOISOTOPES; BODY; CAMERAS; CENTRAL NERVOUS SYSTEM; COUNTING TECHNIQUES; DISEASES; GALLIUM ISOTOPES; GAMMA CAMERAS; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; ISOTOPES; MEASURING INSTRUMENTS; NERVOUS SYSTEM; NUCLEI; ODD-ODD NUCLEI; ORGANS; PROPORTIONAL COUNTERS; RADIATION DETECTORS; RADIOISOTOPES
Optional Information
- Notes
- Updated automatically by Metadata and Full-Text Enrichment Agent