Analysis of thermal conductivity probes for high temperature applications

Description

Various analyses performed to confirm the viability and potential accuracy of a thermal conductivity probe apparatus are summarized and conclusions are presented. The apparatus under consideration was designed for measurements of conductivity of UO2 powders in inert atmospheres at elevated temperatures to 2300 K in support of the Post Accident Heat Removal (PAHR) program at Sandia. Both analytical solutions and numerical computer models are used for study of the apparatus design. The effects of probe thermal mass and limited sample size on measurement accuracy are analyzed and discussed. Results indicate that the designs considered have the potential for accuracies within 5 percent in the conductivity range of interest

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS., PC A04/MF A01.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Additional titles

Augmented title (English)
LMFBR

Publishing Information

Imprint Pagination
52 p.
Report number
SAND--81-0106