Description

The Resistive Plate Chambers (RPC) are used for muon triggers in the CMS experiment. To calibrate the high voltage working-points (WP) and identify degraded detectors due to radiation or chemical damage, a high voltage scan has been performed using 2017 data from pp collisions at a center-of-mass energy of 13 TeV. In this paper, we present the calibration method and the latest results obtained for the 2017 data. A comparison with all scans taken since 2011 is considered to investigate the stability of the detector performance in time.

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/14/09/C09046

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
14
Journal Issue
09
Journal Page Range
p. C09046
ISSN