Published April 1992 | Version v1
Journal article

A low-cost multichannel analyzer with data reduction assembly for continuous air monitoring system

  • 1. Dept. of Electronics Engineering Technology, Texas A and M Univ., College Station, TX (United States)
  • 2. Los Alamos National Lab., Los Alamos, NM (United States)

Description

This paper reports on a microcontroller-based multichannel analyzer (MCA) with a data reduction assembly (DRA) for a plutonium continuous air monitor (CAM) system. The MCA is capable of detecting the airborne alpha emitters in the presence of radon daughter products. The pulse output from the preamplifier has been stretched to allow the peak detector sufficient time to capture the pulse height. The pulse amplitude conversion, the data acquisition, and the output functions are carried out fully by software. The DRA consists of a data reduction unit (DRU) and its operator interface panel. The data reduction assembly has the ability to be networked to a single PC with up to 332 different CAM's remotely connected to it

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
39
Journal Issue
2
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
299-302
ISSN
0018-9499
CODEN
IETNA