Published December 1, 2018
| Version v1
Journal article
Optical frequency measurement comparison using fiber laser combs between NIMT and CMS
Creators
- 1. Department of Dimensional Metrology, National Institute of Metrology (Thailand)
- 2. Center for Measurement Standards, Industrial Technology Research Institute (Thailand)
Description
A NIMT optical frequency comb was compare with an Er-fiber laser comb with repetition frequency of 250 MHz made by the Center for Measurement Standards (CMS), Taiwan by simultaneously measuring the frequency of a 633 nm He-Ne laser. The difference of average frequency measurement is 283 Hz and corresponds to a relative difference of 6.0x10-13. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1742-6596/1144/1/012185Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 1144
- Journal Issue
- 1
- Journal Page Range
- [4 p.]
- ISSN
- 1742-6596
Conference
- Title
- Siam Physics Congress 2018 on a Creative Path to Sustainable Innovation
- Acronym
- SPC2018
- Dates
- 21-23 May 2018
- Place
- Pitsanulok (Thailand)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 53035734
- Subject category
- S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- FIBERS; FREQUENCY MEASUREMENT; HELIUM-NEON LASERS; MHZ RANGE
- Descriptors DEC
- FREQUENCY RANGE; GAS LASERS; LASERS