Published October 1982
| Version v1
Journal article
Modifications of the HEX program for fast automatic resolution of PIXE-spectra
Description
The computer program 'HEX' for resolving PIXE-spectra has been modified. By irradiating standard foils, the escape peak ratios for x-ray peaks in the energy interval 3.3-10.6 keV have been measured. The results have been compared with theoretical calculations and experimental results of other research groups. The escape peaks have been included in the computer code. Pile-up peaks in the x-ray spectrum are also fitted by the modified computer program. This procedure is independent of the count rate and the pile-up interval. In addition, an effective routine for rejecting small x-ray peaks has been developed. (author)
Additional details
Publishing Information
- Journal Title
- X-Ray Spectrom., XRS
- Journal Volume
- 11
- Journal Issue
- 4
- Series
- X-Ray Spectrom., XRS.
- Journal Page Range
- 194-200
- ISSN
- 0049-8246
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 14749054
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- ABSORPTION SPECTRA; AEROSOLS; ALGORITHMS; COMPARATIVE EVALUATIONS; COMPUTER CALCULATIONS; COMPUTER CODES; ESCAPE PEAKS; FOILS; GAUSS FUNCTION; KEV RANGE 01-10; KEV RANGE 10-100; MASS SPECTROSCOPY; PIXE ANALYSIS; X-RAY SPECTRA