Published March 2016
| Version v1
Journal article
Experimental verification of mathematical efficiency modeling and true coincidence summing correction of the SAGe Well detector
- 1. Canberra Industries Inc., Meriden, CT (United States)
- 2. Stroiteley Street 4-4-52, Moscow (Russian Federation)
Description
The characterization and ray-tracing algorithm in ISOCS/LabSOCS has been extended to calculate full energy peak and total efficiencies for the Small Anode Germanium Well detector. The extension allows for true coincidence summing correction using the standard algorithm in the Genie 2000 software. The accuracy of the full energy peak efficiency calculation and the coincidence summing correction has been experimentally validated to be the same as for non-well detectors. (author)
Additional details
Publishing Information
- Journal Title
- Journal of Radioanalytical and Nuclear Chemistry
- Journal Volume
- 307
- Journal Issue
- 3
- Journal Page Range
- p. 1961-1966
- ISSN
- 0236-5731
- CODEN
- JRNCDM
Conference
- Title
- 10. International Conference on Methods and Applications of Radioanalytical Chemistry
- Acronym
- MARC X
- Dates
- 12-17 Apr 2015
- Place
- Kailua-Kona (United States)
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 47051343
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ANODES; COINCIDENCE METHODS; COMPUTERIZED SIMULATION; GE SEMICONDUCTOR DETECTORS; MATHEMATICAL MODELS; QUANTUM WELLS
- Descriptors DEC
- COUNTING TECHNIQUES; ELECTRODES; MEASURING INSTRUMENTS; NANOSTRUCTURES; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SIMULATION
Optional Information
- Notes
- 18 refs.