Published 1979
| Version v1
Report
Restricted
Practical measuring facilities for direct incorporation monitoring
Description
A standardized monitoring system is presented consisting of a shadow-shield partial-body monitor and a thyroid monitor with detector shielding. Both of the monitors contain NaI(Tl) detectors. With 63 mm diameter x 63 mm thick crystals, the minimum detectable activities for a 5-minute measurement were found to be about 370 Bq (10 nCi) of 60Co in the lungs and 111 Bq (3 nCi) of 131I in the thyroid. (author)
Availability note (English)
MF available from INIS under the Report Number.
Files
Additional details
Additional titles
- Original title (German)
- Praktische Messeinrichtungen fuer direkte Inkorporationskontrollen
Publishing Information
- Imprint Pagination
- 27 p.
- Report number
- SAAS--247
INIS
- Country of Publication
- German Democratic Republic
- Country of Input or Organization
- German Democratic Republic
- INIS RN
- 10487641
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- CALIBRATION; COBALT 60; DOSIMETRY; IODINE 125; IODINE 131; LUNGS; PHANTOMS; RADIATION MONITORING; RADIONUCLIDE KINETICS; SHIELDING; SHIELDING MATERIALS; SOLID SCINTILLATION DETECTORS; THYROID
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BODY; COBALT ISOTOPES; DAYS LIVING RADIOISOTOPES; ELECTRON CAPTURE RADIOISOTOPES; ENDOCRINE GLANDS; GLANDS; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTO; IODINE ISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; KINETICS; MEASURING INSTRUMENTS; MINUTES LIVING RADIOISOTOPES; MOCKUP; MONITORING; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; ORGANS; RADIATION DETECTORS; RADIOISOTOPES; RESPIRATORY SYSTEM; SCINTILLATION COUNTERS; STRUCTURAL MODELS; YEARS LIVING RADIOISOTOPES