Published November 2001 | Version v1
Journal article

A discretized approach to determining TG-43 brachytherapy dosimetry parameters: case study using Monte Carlo calculations for the MED3633 103Pd source

Creators

Description

It is of interest to discern the energy-dependence of American Association of Physicists in Medicine (AAPM) TG-43 brachytherapy dosimetry parameters. Using Monte Carlo calculation geometry and techniques (MCNP), dependence of these parameters was calculated as a function of photon energy, in general, and for the MED3633 103Pd source using a discretized approach. Results were weighted and summed to determine the total contribution for comparison with the 103Pd source literature. Comprehensive 2-D results are discussed, and the level of agreement with other assessments are presented

Additional details

Identifiers

PII
S0969804301001440;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
55
Journal Issue
6
Journal Page Range
p. 775-782
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Copyright
Copyright (c) 2001 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.