Accuracy of amplitude-based respiratory gating for PET/CT in irregular respirations
Creators
- 1. Kyushu University Hospital, Department of Medical Technology, Fukuoka (Japan)
- 2. Kyushu University, Graduate School of Medical Sciences, Fukuoka (Japan)
Description
We evaluated the accuracy of amplitude gating PET (AG-PET) compared with phase gating PET (PG-PET) in relation to respiratory motion patterns based on a phantom analysis. We used a NEMA IEC body phantom filled with an 18F solution with a 4:1 sphere-to-background radioactivity ratio (12.6 and 2.97 kBq/mL). PET/CT scans were acquired in a motionless and moving state on a Biograph mCT. The respiratory movements were simulated by four different waveform patterns consisting of ideal breathing, breathing with a pause period, breathing with a variable amplitude and breathing with a changing baseline. AG-PET selects the narrow bandwidth containing 20% of the respiratory cycle. PG-PET was reconstructed with five gates. The image quality was physically assessed using the percent contrast (QH,10mm), background variability (N10mm) recovery coefficient (RC), and sphere volumes. In regular motion patterns with ideal breathing and breathing with a pause period, the QH,10mm RC and sphere volumes were not different between AG-PET and PG-PET. In the variable amplitude pattern, the QH,10mm of AG-PET was higher than that of PG-PET (35.8 vs 28.2%), the RC of AG-PET was higher than that of PG-PET and sphere volume of AG-PET was smaller than that of PG-PET (6.4 vs 8.6 mL). In the changing baseline pattern, the QH,10mm of AG-PET was higher than that of PG-PET (42.4 vs 16.7%), the RC of AG-PET was higher than that of PG-PET and sphere volume of AG-PET was smaller than that of PG-PET (6.2 vs 9.8 mL). The N10mm did not differ between AG-PET and PG-PET, irrespective of the motion pattern. Amplitude gating PET is considered to be more accurate than phase gating PET for examining unstable respiratory motion patterns, such as those involving a variable amplitude or changing baseline. (author)
Availability note (English)
Available from http://dx.doi.org/10.1007/s12149-014-0870-5Additional details
Identifiers
Publishing Information
- Journal Title
- Annals of Nuclear Medicine
- Journal Volume
- 28
- Journal Issue
- 8
- Journal Page Range
- p. 770-779
- ISSN
- 0914-7187
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 46090069
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; AMPLITUDES; COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; MOTION; PHANTOMS; POSITRON COMPUTED TOMOGRAPHY; RESPIRATION; SIMULATION; SYNCHRONIZATION
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; EVALUATION; MOCKUP; STRUCTURAL MODELS; TOMOGRAPHY