Contribution to the study and to the development of continuous infrared absorption analyzers

  • 1. Commissariat a l'Energie Atomique, Saclay (France). Centre d'Etudes Nucleaires

Description

The continuous infra-red absorption determination of corrosive gases implies the construction of an original single-beam spectrometer. The use of a vibration-rotation line means that the apparatus function has to undergo an alternating displacement: for one position this function is centered on the line, for other position it is placed close by, but away from any absorption. A non-linear relationship is derived between the concentration of the impurity under consideration, the optical density measured at the centre of the line and the total pressure of the gas mixture. The problem is then solved using an associated specific analog computer. The system is applicable to the determination of low concentrations. The principle proposed in then adapted to an analysis using a non-resolved vibration-rotation band. (authors)

Availability note (English)

Available from INIS in electronic form

Abstract (French)

Le dosage en continu par absorption infrarouge des gaz corrosifs, implique la construction d'un spectrometre original en simple faisceau. L'utilisation d'une raie de vibration - rotation, conduit a effectuer un deplacement alternatif de la fonction d'appareil: pour une position, celle-ci est centree sur la raie, pour l'autre position elle est placee a proximite mais en dehors de toute absorption. On etablit une relation non lineaire entre la concentration de l'impurete consideree, la densite optique mesuree au centre de la raie et la pression totale du melange gazeux. Le probleme est alors resolu par un calculateur analogique specifique associe. Le systeme s'applique au dosage des faibles concentrations. Le principe propose est ensuite adapte a l'analyse a partir d'une bande de vibration - rotation non resolue. (auteurs)

Files

35091367.pdf

Files (2.7 MB)

Name Size Download all
md5:0b5eac0d2ab7daf4ac85f8058c81d174
2.7 MB Preview Download

Additional details

Additional titles

Original title (French)
Contribution a l'etude et a la mise au point des analyseurs en continu par absorption infrarouge

Publishing Information

Imprint Pagination
[108 p.]
Report number
CEA-R--3527

Optional Information

Notes
45 refs.