Published December 1, 1989 | Version v1
Journal article

A 'TDPAC-camera'

  • 1. Technische Univ. Muenchen, Garching (Germany, F.R.). Fakultaet fuer Physik

Description

An advanced spectrometer for time differential perturbed γ-γ-angular correlation measurements is presented. It consists of six detectors with BaF2-scintillators and is designed to reduce the data collection time as much as possible. Experimental results with 'illumination' times of the order of 100 s are presented. This 'camera' storing 30 coincidence subgroups simultaneously is compared with a conventional four-detector system with eight subgroups. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research, Section A
Journal Volume
284
Journal Issue
2/3
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
417-421
ISSN
0168-9002
CODEN
NIMAE