To calculate a safe and effective 153Sm-EDTMP therapy dose, a whole-body scintigraphy technique for prospective individual dosimetry was developed and the results were compared with 5 h urine collection method in 20 patients with bone metastases. Anterior and posterior whole-body images were obtained 10 min and 5 h after intravenous injection of 740 MBq 153Sm-EDTMP and the bone uptake value was determined. There is a close correlation between the bone uptake value obtained from the whole-body scintigraphy and 5 h urine collection method (r = 0.93). The radiation absorption dose to red marrow was limited to 1.4 Gy and the administered activity calculated from bone uptake value by whole-body scintigraphy was 1.40-2.27 GBq (mean 1.90 GBq). If the activity was calculated according to the standard body weight of 37 MBq·kg-1, the administered activity would be 1.75-2.41 GBq (mean 2.18), the radiation absorption dose to red marrow would be 1.37-2.27 Gy (mean 1.63 Gy), but at these doses significant myelotoxicity would be anticipated, thus emphasizing the need for individual prospective dosimetry

Additional details

Publishing Information

Journal Title
Nuclear Science and Techniques
Journal Volume
11
Journal Issue
4
Journal Page Range
p. 257-262
ISSN
1001-8042