Published July 1987
| Version v1
Journal article
Characteristics of hodoscopic spectromagnetic calorimeter from lead glass with 2 m2 aperture
- 1. Joint Inst. for Nuclear Research, Dubna (USSR)
Description
Cherenkov lead-glass-base electromagnetic calorimeter, comprising a shower hodoscopic detector with 10x10 cm dimension cell (189 elements) and active converter with the thickness of 4 radiation lengths (19 elements), mounted in front of it, is described. The detector energy resolution of ∼ 0.08/√E (E is measured in gigaelectronvolts), the accuracy of reconstructing coordinates of γ-quanta and electrons (positrons) of ∼ 10 mm, the level of discrimination of signals from hadrons of ∼ 2x10-3 are measured under irradiation in beams of positrons and π+-mesons with 3 and 5 GeV energies
Additional details
Additional titles
- Original title (Russian)
- Характеристики годоскопического электромагнитного калориметра из свинцового стекла с апертурой 2 м
Publishing Information
- Journal Title
- Prib. Tekh. Ehksp.
- Journal Issue
- no.4
- Series
- Prib. Tekh. Ehksp.
- ISSN
- 0032-8162
- CODEN
- PRTEA
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- USSR
- INIS RN
- 19046722
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKGROUND NOISE; CALORIMETERS; CHERENKOV COUNTERS; DIAGRAMS; ENERGY RESOLUTION; GEV RANGE 01-10; HODOSCOPES; LEAD OXIDES; PERFORMANCE TESTING; SHOWER COUNTERS; SPATIAL RESOLUTION; STABILITY
- Descriptors DEC
- CHALCOGENIDES; ENERGY RANGE; GEV RANGE; INFORMATION; LEAD COMPOUNDS; MEASURING INSTRUMENTS; NOISE; OXIDES; OXYGEN COMPOUNDS; RADIATION DETECTORS; RESOLUTION; TESTING