Published March 2001 | Version v1
Journal article

Position sensitive detector for polarized neutron spectrometer

  • 1. Ob''edinennyj Inst. Yadernykh Issledovanij, Dubna (Joint Institute for Nuclear Research (JINR))
  • 2. Laue-Langevin Inst., Grenoble (France)

Description

Thermal neutron linear detector designed in the JINR Laboratory of High Energies is peculiar for the JINR Ibr-2 reactor polarized neutron spectrometer to record thermal neutrons. Information is accumulated in the course of frame-by-frame photography to discern neutrons with λ = 1-12 A wave length as to time of flight. The efficiency of neutron recording constitutes about 70% for 2 A wave length. The detector is low sensitive to γ-radiation, is of about 1.5% differential nonlinearity and is of about 1.5 mm space resolution at about 105 events per second counting rates

Abstract (Russian)

Разработанный в Лаборатории высоких энергий ОИЯИ линейный детектор тепловых нейтронов предназначен для регистрации тепловых нейтронов на спектрометре поляризованных нейтронов реактора ИБР-2 ОИЯИ. Набор информации ведется в режиме кадровой съемки для выделения нейтронов с длиной волны λ = 1-12 А по времени пролета. Эффективность регистрации нейтронов составляет ∼ 70 % для длины волны 2 А. Детектор имеет низкую чувствительность к γ-излучению, дифференциальную нелинейность ∼ 1.5% и пространственное разрешение ∼ 1.5 мм при скоростях счета ∼ 105 событий за секунду

Additional details

Additional titles

Original title (Russian)
Позиционно-чувствительный детектор для спектрометра поляризованных нейтронов

Publishing Information

Journal Title
Pribory i Tekhnika Ehksperimenta
Journal Issue
no.2
Journal Page Range
p. 5-12
ISSN
0032-8162
CODEN
PRTEAJ

Optional Information

Notes
7 refs., 8 figs.