Published 2010 | Version v1
Miscellaneous Open

Implantation of total body irradiation in radiotherapy

Description

Before implementing a treatment technique, the characteristics of the beam under irradiation conditions must be well acknowledged and studied. Each one of the parameters used to calculate the dose has to be measured and validated before its utilization in clinical practice. This is particularly necessary when dealing with special techniques. In this work, all necessary parameters and measurements are described for the total body irradiation implementation in facilities designed for conventional treatments that make use of unconventional geometries to generate desired enlarged field sizes. Furthermore, this work presents commissioning data of this modality at Hospital das Clinicas of Sao Paulo using comparison of three detectors types for measurements of entrance dose during total body irradiation treatment. (author)

Files

43130556.pdf

Files (1.3 MB)

Name Size Download all
md5:b2df66a8ec6e4e3ba6510da0a38f1184
1.3 MB Preview Download

Additional details

Additional titles

Original title (Portuguese)
Implementacao da irradiacao de corpo inteiro em radioterapia

Publishing Information

Imprint Pagination
59 p.