Published May 1, 2008 | Version v1
Journal article

A fast response cadmium-selective polymeric membrane electrode based on N,N'-(4-methyl-1,2-phenylene)diquinoline-2-carboxamide as a new neutral carrier

  • 1. Department of Chemistry, Isfahan University of Technology, Isfahan 84156-84111 (Iran, Islamic Republic of)

Description

A new polyvinylchloride membrane sensor for Cd2+ ions based on N,N'-(4-methyl-1,2-phenylene)diquinoline-2-carboxamide (Mebqb) as a new and excellent neutral ionophore has been prepared. The sensor shows a Nernestian response for cadmium ions over a wide concentration range (1.0 x 10-6 to 1.0 x 10-1 M) with the determination coefficient of 0.9964 and slope of 29.9 ± 0.5 mV decade-1. The limit of detection is 8 x 10-7 M. It has a fast response time of 3-8 s and can be used for at least 8 weeks without any divergence in potential. The electrode can be used in the pH range from 4.0 to 9.0. The proposed sensor shows a very good discriminating ability towards Cd2+ ion in comparison to some alkali, alkaline earth, transition and heavy metal ions. It was successfully applied for the direct determination of Cd2+in standard and real sample solutions

Availability note (English)

Available from http://dx.doi.org/10.1016/j.jhazmat.2007.08.033

Additional details

Identifiers

DOI
10.1016/j.jhazmat.2007.08.033;
PII
S0304-3894(07)01210-1;

Publishing Information

Journal Title
Journal of Hazardous Materials
Journal Volume
153
Journal Issue
1-2
Journal Page Range
p. 179-186
ISSN
0304-3894
CODEN
JHMAD9

Optional Information

Copyright
Copyright (c) 2007 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.