Off-axis beam quality change in linear accelerator x-ray beams

  • 1. Radiological Physics Center, The University of Texas System Center Center M. D. Anderson Hospital and Tumor Institute, Houston, Texas 77025

Description

The effective energy of the x-ray beam from linear accelerators changes as a function of the position in the beam due to nonuniform filtration by the flattening filter. In this work, the transmittance through a water column was measured in good geometry and the beam quality characterized in units of HVL in water. Measurements were made on a variety of linear accelerators from 4 to 10 MV. The beam energy decreased with increasing distance from the central ray for all accelerators measured