Calibration of low-temperature ac susceptometers with a copper cylinder standard
Creators
- 1. ICREA and Departament de Fisica, Universitat Autonoma de Barcelona, Bellaterra, 08193 Barcelona (Spain)
Description
A high-quality low-temperature ac susceptometer is calibrated by comparing the measured ac susceptibility of a copper cylinder with its eddy-current ac susceptibility accurately calculated. Different from conventional calibration techniques that compare the measured results with the known property of a standard sample at certain fixed temperature T, field amplitude Hm, and frequency f, to get a magnitude correction factor, here, the electromagnetic properties of the copper cylinder are unknown and are determined during the calibration of the ac susceptometer in the entire T, Hm, and f range. It is shown that the maximum magnitude error and the maximum phase error of the susceptometer are less than 0.7% and 0.3 deg., respectively, in the region T=5-300 K and f=111-1111 Hz at Hm=800 A/m, after a magnitude correction by a constant factor as done in a conventional calibration. However, the magnitude and phase errors can reach 2% and 4.3 deg. at 10 000 and 11 Hz, respectively. Since the errors are reproducible, a large portion of them may be further corrected after a calibration, the procedure for which is given. Conceptual discussions concerning the error sources, comparison with other calibration methods, and applications of ac susceptibility techniques are presented.
Additional details
Identifiers
- DOI
- 10.1063/1.3309779;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 81
- Journal Issue
- 2
- Journal Page Range
- p. 025104-025104.10
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44013608
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; CALIBRATION STANDARDS; COMPARATIVE EVALUATIONS; COPPER; CORRECTIONS; CYLINDERS; EDDY CURRENTS; ERRORS; MAGNETIC SUSCEPTIBILITY; TEMPERATURE RANGE 0065-0273 K
- Descriptors DEC
- CURRENTS; ELECTRIC CURRENTS; ELEMENTS; EVALUATION; MAGNETIC PROPERTIES; METALS; PHYSICAL PROPERTIES; STANDARDS; TEMPERATURE RANGE; TRANSITION ELEMENTS
Optional Information
- Notes
- (c) 2010 American Institute of Physics