Published August 1996 | Version v1
Report Open

A program on quality assurance and dose calibration for radiation therapy units in Venezuela

  • 1. Instituto Venezolano de Investigaciones Cientificas, Caracas (Venezuela). Lab. de Calibracion Dosimetrica

Description

The results of a five year program (1988-90-91-92-93) on quality assurance and dose calibration for 12 Cobalt-60 units from public hospitals, which represents 30% of total radiation therapy units in Venezuela, are presented. The remarkable improvement in the general performance of these units can be seen in the IAEA/WHO Postal TLD Intercomparison results which gave 100% within ± 5% in 1990 and 1992, while 63% in 1990 and 44% in 1992, with errors up to 37% were obtained for the participants not included in the program. The difference between the two groups lead the government to decrete through the Gaceta Oficial de la Republica de Venezuela, Resolution G-1397 on March 3, 1993, the quality assurance and dose calibration programs shall be established for all radiation therapy installations in Venezuela. The project for the standards was developed by the SSDL physicists and it was already approbated by the Health Ministry. It is expected that the Norms will enter into effect by the end of 1994. (author). 14 refs, 1 fig., 3 tabs

Files

28018212.pdf

Files (205.1 kB)

Name Size Download all
md5:05e982ed91d7f980c2ae2d8fbbbf08d8
205.1 kB Preview Download
Part of:
Radiation dose in radiotherapy from prescription to delivery

Additional details

Publishing Information

Imprint Title
Radiation dose in radiotherapy from prescription to delivery
Imprint Pagination
293 p.
Journal Page Range
p. 135-140.
ISSN
1011-4289
Report number
IAEA-TECDOC--896

Conference

Title
Radiation dose in radiotherapy from prescription to delivery.
Acronym
Interregional seminar on radiotherapy dosimetry
Dates
27-30 Aug 1994.
Place
Rio de Janeiro (Brazil).