Beam test results of a PbWO4 crystal calorimeter prototype
Creators
- 1. Institut Fiziki Vysokikh Ehnergij, Protvino (Russian Federation)
- 2. Universite Libre de Bruxelles, CP 229, 1050 Brussels (Belgium)
- 3. Institute of Nuclear Problems, 220050 Minsk (Belarus)
- 4. Laboratoire d'Annecy de Physique des Particules, BP 110, 74941 Annecy-le-Vieux (France)
Description
Beam tests of an EM-calorimeter prototype made of PbWO4 crystals, produced after significant improvement in growth technology, have been performed at the CERN SPS. The measured energy resolution is σE/E=2.8%/√E+0.47%. (orig.)
Additional details
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 364
- Journal Issue
- 2
- Journal Page Range
- p. 307-310.
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 27002504
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALORIMETERS; CRYSTALS; ELECTRON DETECTION; ENERGY DEPENDENCE; ENERGY RESOLUTION; GEV RANGE 10-100; LEAD TUNGSTATES; RELATIVISTIC RANGE; SOLID SCINTILLATION DETECTORS; TEMPERATURE DEPENDENCE; TEMPERATURE RANGE 0273-0400 K
- Descriptors DEC
- CHARGED PARTICLE DETECTION; DETECTION; ENERGY RANGE; GEV RANGE; LEAD COMPOUNDS; MEASURING INSTRUMENTS; OXYGEN COMPOUNDS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION; SCINTILLATION COUNTERS; TEMPERATURE RANGE; TRANSITION ELEMENT COMPOUNDS; TUNGSTATES; TUNGSTEN COMPOUNDS