Published May 1999 | Version v1
Journal article

Criteria for designing an MRI quality assurance program. A multicenter trial

  • 1. Brescia Spedali Civili, Brescia (Italy). Servizio di fisica sanitaria
  • 2. Varese Ospedale di Circolo, Varese (Italy). Servizio di fisica sanitaria
  • 3. Milan Ospedale Ca' Granda Niguarda, Milan (Italy). Servizio di fisica sanitaria
  • 4. Milan Ospedale S. Carlo, Milan (Italy)
  • 5. Milan Ospedale S. Raffaele, Milan (Italy). Servizio ingegneria biomedica

Description

The authors report the preliminary results of a multicenter trial aimed at defining methods, reference values and frequency of measurements for a magnetic resonance quality assurance program. In particular, they stress the definition of two attention levels (investigation and intervention) for image uniformity and signal-to-noise ratio (SNR) by means short-and long-term measurements

Abstract (Italian)

Il lavoro presenta i risultati preliminari di uno studio multicentrico per individuare metodi, valori di riferimento e frequenze delle misure di un programma del controllo della qualita' dell'immagine nella risonanza magnetica. In particolare propone la definizione di due intervalli di attenzione (indagine e intervento) per il rapporto segnale-rumore e l'uniformita' integrale percentuale, mediante misure a breve e lungo termine

Additional details

Additional titles

Original title (Italian)
Criteri per la definizione di un programma per l'assicurazione della qualita' nella diagnostica per immagini con risonanza magnetica: studio multicentrico

Publishing Information

Journal Title
Radiologia Medica
Journal Volume
97
Journal Issue
5
Journal Page Range
p. 389-397
ISSN
0033-8362
CODEN
RAMEAN

INIS

Country of Publication
Italy
Country of Input or Organization
Italy
INIS RN
31021561
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
DIAGNOSTIC TECHNIQUES; IMAGE SCANNERS; MAGNETIC RESONANCE; PERFORMANCE; QUALITY ASSURANCE
Descriptors DEC
RESONANCE