Published October 2004 | Version v1
Journal article

Charge exchange spectroscopy by Fabry-Perot spectrometer in W7-AS

  • 1. MaxPlank Institute for Plasma Physics, Wendelsteinstrasse, 1, 17491 Greifswald (Germany)
  • 2. National Institute for Fusion Science, Toki 509-5292 (Japan)

Description

Charge exchange spectroscopy using a Fabry-Perot spectrometer has been developed to study the dynamic of ion temperature and radial electric field in plasmas. A charge coupled device detecter with 80x80 pixels was used to gain the spectral resolution of the charge exchange spectroscopy system. This Fabry-Perot charge exchange spectroscopy system has been applied to measure the ion temperature using the charge exchange line of carbon impurity with a time resolution of 5 ms for high density quasisteady discharges in W7-AS. The cold component due to the charge exchange reaction between the carbon impurity and thermal neutrals is subtracted from the change emission with the beam modulation technique

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
75
Journal Issue
10
Journal Page Range
p. 4136-4138
ISSN
0034-6748
CODEN
RSINAK

Conference

Title
15. topical conference on high temperature plasma diagnostics
Dates
19-22 Apr 2004
Place
San Diego, CA (United States)

Optional Information

Notes
(c) 2004 American Institute of Physics