Published September 21, 2015 | Version v1
Journal article

A prototype High Purity Germanium detector for high resolution gamma-ray spectroscopy at high count rates

  • 1. Lawrence Berkeley National Laboratory, 1 Cyclotron Road, Berkeley, CA 94720 (United States)
  • 2. Department of Nuclear Engineering, University of California, Berkeley, CA 94720 (United States)

Description

Where energy resolution is paramount, High Purity Germanium (HPGe) detectors continue to provide the optimum solution for gamma-ray detection and spectroscopy. Conventional large-volume HPGe detectors are typically limited to count rates on the order of ten thousand counts per second, however, limiting their effectiveness for high count rate applications. To address this limitation, we have developed a novel prototype HPGe detector designed to be capable of achieving fine energy resolution and high event throughput at count rates in excess of one million counts per second. We report here on the concept, design, and initial performance of the first prototype device

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2015.05.053

Additional details

Identifiers

DOI
10.1016/j.nima.2015.05.053;
PII
S0168-9002(15)00712-3;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
795
Journal Page Range
p. 167-173
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2015 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.