Development of gamma scintillation camera
- 1. Bhabha Atomic Research Centre, Bombay (India). Electronics Div.
Description
This paper describes the indigenous development of gamma ray scintillation camera in Electronics Division, BARC. The system is based on Anger principle. To establish the know-how the gamma camera was built around a 260 mm dia 6 mm thick NaI (T1) coupled to nineteen 38 mm dia photomultiplier tubes. The system after thorough testing and laboratory evaluation, has been upgraded to a system with thirty seven photomultipliers. (author). 3 figs
Additional details
Publishing Information
- Publisher
- Bhabha Atomic Research Centre.
- Imprint Place
- Bombay (India)
- Imprint Title
- Proceedings of the national symposium on nuclear electronics and instrumentation (held at Bombay during 15-17 February 1989)
- Imprint Pagination
- 662 p.
- Journal Page Range
- p. 341-348.
Conference
- Title
- National symposium on nuclear electronics and instrumentation.
- Dates
- 15-17 Feb 1989.
- Place
- Bombay (India).
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 21072586
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BARC; COLLIMATORS; DIAGNOSIS; ELECTRONIC CIRCUITS; FABRICATION; GAMMA CAMERAS; MONOCRYSTALS; NAI DETECTORS; NUCLEAR MEDICINE; PHOTOMULTIPLIERS; SODIUM IODIDES
- Descriptors DEC
- ALKALI METAL COMPOUNDS; CAMERAS; CRYSTALS; HALIDES; HALOGEN COMPOUNDS; INDIAN ORGANIZATIONS; IODIDES; IODINE COMPOUNDS; MEASURING INSTRUMENTS; MEDICINE; NATIONAL ORGANIZATIONS; PHOTOTUBES; RADIATION DETECTORS; SCINTILLATION COUNTERS; SODIUM COMPOUNDS; SOLID SCINTILLATION DETECTORS