Background considerations for low rate SSG detector experiments
Description
The low energy thresholds achieved with detectors made of superheated superconducting granules (SSG) encourage to use these devices for low rate measurements like WIMP search or coherent neutrino scattering experiments. Such applications need powerful rejection criteria for background events from electronic noise, cosmic and gamma radiation. In order to develop background rejection strategies, the origin of background sources and their interactions with SSG detectors need to be known. We present a discussion about electronic noise as well as background radiation together with their effects on the SSG detector counting rate. (orig.)
Additional details
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 374
- Journal Issue
- 1
- Journal Page Range
- p. 111-117.
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 27060098
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKGROUND NOISE; BACKGROUND RADIATION; COSMIC MUONS; COUNTING RATES; EFFICIENCY; GAMMA RADIATION; KEV RANGE 100-1000; LOW LEVEL COUNTING; MEV RANGE 01-10; PARTICLE DISCRIMINATION; RADIATION DETECTION; RADIATION DETECTORS; SUPERCONDUCTING DEVICES; TEMPERATURE RANGE 0000-0013 K; TIN
- Descriptors DEC
- COSMIC RADIATION; COUNTING TECHNIQUES; DETECTION; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; ELEMENTS; ENERGY RANGE; FERMIONS; IONIZING RADIATIONS; KEV RANGE; LEPTONS; MEASURING INSTRUMENTS; METALS; MEV RANGE; MUONS; NOISE; PARTICLE IDENTIFICATION; RADIATIONS; SECONDARY COSMIC RADIATION; TEMPERATURE RANGE