Published March 2017 | Version v1
Journal article

Radionuclide dosimetry with radiochromic gels: a feasibility study

  • 1. Czech Metrology Institute, Brno(Czech Republic)

Description

Radiochromic gel dosimeters were studied with respect to their potential as tools to validate Monte Carlo models and procedures used in molecular radiotherapy (MRT). The Fricke xylenol orange gel (FX gel) and the Turnbull Blue dye gel (TB gel) containing typical carriers of selected radionuclides used in MRT were examined. 177Lu was the only radionuclide that did not impair the optical, chemical or dosimetric properties of the FX gel. The TB gel was found unusable altogether because all of the carriers tested either brought about a significant increase in the gel's spontaneous background or caused gel precipitation. The difference between the absorbed dose-to-water observed by using the FX gel with 177Lu and the results of the Monte Carlo simulation using the MCNPX™ code was about 10%. However, the measurement uncertainty was so high that it precluded the use of the gel for validation of the Monte Carlo results.

Additional details

Publishing Information

Journal Title
Bezpecnost Jaderne Energie
Journal Volume
25
Journal Issue
3-4
Journal Page Range
p. 119-128
ISSN
1210-7085

Conference

Title
15. St. Nicholas meeting of the Young Generation of the Czech Nuclear Society
Original Conference Title
15. mikulášské setkání Mladé generace České nukleární společnosti
Dates
2-4 Dec 2015
Place
Brno (Czech Republic)

Optional Information

Notes
5 tabs., 5 figs., 25 refs.