Simultaneous determination of copper, lead and cadmium in tantalum by square-wave polarography
Creators
- 1. Nagoya Univ. (Japan). Faculty of Engineering
Description
Copper, lead and cadmium at the ppb level in high-purity tantalum metal are determined simultaneously by square-wave polarography after removal of tantalum by anion exchange. A 1-g sample is dissolved in a mixture of hydrofluoric and nitric acids in a Teflon beaker, and the solution (20 ml, 4 M in F- and 0.5 M in NO3-) is introduced onto an anion-exchange column (Bio·Rad AG 1-X8, 100200 mesh, 10 mmφ × 150 mm, polyethylene tubing). Then 20 ml of the mixed acid (4 M HF +0.5 M HNO3) is passed through the column. The effluent is evaporated to dryness and the residue is dissolved in 5 ml of the supporting electrolyte containing 0.5 M potassium chloride-0.1 M hydrochloric acid. The lower limits of determination are 0.05 ppm for copper, 0.1 ppm for lead, and 0.02 ppm for cadmium, and the precision is about 10% at the 0.2 ppm level. The time required for a determination is approximately 5 to 6 hours.
Additional details
Identifiers
Publishing Information
- Journal Title
- BUNSEKI KAGAKU
- Journal Volume
- 21
- Journal Issue
- 12
- Series
- Bunseki Kagaku.
- Journal Page Range
- 1645-1647
- ISSN
- 0525-1931
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 4053047
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- No Abstract
- Descriptors DEI
- ACCURACY; CADMIUM; COPPER; ELECTROLYTES; ION EXCHANGE CHROMATOGRAPHY; LEAD; POLAROGRAPHY; QUANTITATIVE CHEMICAL ANALYSIS; TANTALUM
- Descriptors DEC
- CHEMICAL ANALYSIS; CHROMATOGRAPHY; ELEMENTS; METALS; SEPARATION PROCESSES; TRANSITION ELEMENTS
Optional Information
- Notes
- Letter-to-the-editor.; Updated automatically by Metadata and Full-Text Enrichment Agent