Robotically Assisted CBCT-Guided Needle Insertions: Preliminary Results in a Phantom Model
Creators
- 1. University of Strasbourg, CNRS, INSA Strasbourg, ICube Laboratory (France)
- 2. University Hospital of Strasbourg, Department of Interventional Radiology (France)
- 3. Queen Marie Hospital, Hong Kong, Interventional Radiology (Hong Kong)
Description
Aim
To compare robotic-assisted needle insertions performed under CBCT guidance to standard manual needle insertions.Materials and Methods
A homemade robotic prototype was used by two operators to perform robotic and manual needle insertions on a custom-made phantom. Both the operators had no experience with the prototype before starting the trial. The primary endpoint was accuracy (i.e., the minimal distance between the needle tip and the center of the target) between robotic and manual insertions. Secondary endpoints included total procedure time and operators' radiation exposure. The Wilcoxon test was used. A p value less than 0.05 was considered statistically significant.
Results
Thirty-three (17 manual, 16 robotic) needle insertions were performed. Mean accuracy for robotic insertion was 2.3 ± 0.9 mm (median 2.1; range 0.8–4.2) versus 2.3 ± 1 mm (median 2.1; range 0.7–4.4) for manual insertion (p = 0.84). Mean procedure time was 683 ± 57 s (median 670; range 611–849) for the robotic group versus 552 ± 40 s (median 548; range 486–621) for the manual group (p = 0.0002). Mean radiation exposure was 3.25 times less for the robotic insertion on comparison to manual insertion for the operator 1 (0.4 vs 1.3 µGy); and 4.15 times less for the operator 2 (1.9 vs 7.9 µGy).
Conclusion
The tested robotic prototype showed accuracy comparable to that achieved with manual punctures coupled to a significant reduction of operators' radiation exposure. Further, in vivo studies are necessary to confirm the efficiency of the system.
Additional details
Identifiers
Publishing Information
- Journal Title
- Cardiovascular and Interventional Radiology
- Journal Volume
- 42
- Journal Issue
- 2
- Journal Page Range
- p. 283-288
- ISSN
- 0174-1551
- CODEN
- CAIRDG
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 51117726
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; BIOPSY; COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; FLUOROSCOPY; MULTIVARIATE ANALYSIS; OCCUPATIONAL EXPOSURE; PHANTOMS; RADIATION DOSES; ROBOTS
- Descriptors DEC
- BIOMEDICAL RADIOGRAPHY; DIAGNOSTIC TECHNIQUES; DOSES; EQUIPMENT; EVALUATION; MATHEMATICS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIOLOGY; STATISTICS; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Copyright
- Copyright (c) 2019 Springer Science+Business Media, LLC, part of Springer Nature and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE)