An intercomparison of several environmental TLD systems

  • 1. Pennsylvania State Univ., University Park (USA). Dept. of Nuclear Engineering

Description

Three environmental dosemeter systems, Panasonic UD-200S, TI, and Panasonic UD-801A provided by two utilities, were evaluated. The absolute system calibrations and the system dependent variations in the derived doses were investigated by irradiating the dosemeters at three different exposure levels in a well characterised photon field (137Cs/137 Basup(m)) in the laboratory. The resulting TL response data were analysed using standard statistical methods. The UD-801A system results, while deviating the most from the delivered doses, exhibit markedly lower fluctuations than the others. This effect appears to be the result of using a stringent statistical selection criterion and determining individual element calibration factors according to the guidelines of Plato and Miklos. Also, the chi squared analysis indicates that the normal distribution model does not provide a highly accurate representation of the data. (author)

Additional details

Publishing Information

Journal Title
Radiat. Prot. Dosim.
Journal Volume
15
Journal Issue
3
Series
Radiat. Prot. Dosim.
Journal Page Range
177-183
ISSN
0144-8420
CODEN
RPDOD

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
18028753
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CALIBRATION; COMPARATIVE EVALUATIONS; ENVIRONMENT; PERFORMANCE TESTING; THERMOLUMINESCENT DOSEMETERS
Descriptors DEC
DOSEMETERS; EVALUATION; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; TESTING