Published October 1997
| Version v1
Report
Open
Research and development of position sensitive silicon detectors
Creators
- 1. Laboratoire de Physique Subatomique et des Technologies Associees - SUBATECH, Centre National de la Recherche Scientifique, 44 - Nantes (France)
- 2. Institut de Physique Nucleaire, Paris-11 Univ., 91 - Orsay (France)
Description
A two-dimensional position sensitive silicon detector was tested. Linearity and resolution of the position were investigated and found to be better than 50 μ m. To estimate the spatial resolution we have used a laser diode mounted on a X-Y table. An example of sweeping of the active area of the detector is presented. Initially, conceived to improve the two-particle correlation measurements, these detectors could be of interest in experiments requiring precise kinematical reconstructions
Files
29040477.pdf
Files
(57.1 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:a44b482e81cde7240a386807ceca4688
|
57.1 kB | Preview Download |
Additional details
Additional titles
- Original title (French)
- Recherche et developpement sur des detecteurs silicium a localisation
Publishing Information
- Imprint Title
- Report of Activity, 1994 - 1996
- Imprint Pagination
- 212 p.
- Journal Page Range
- p. 84
- Report number
- SUBATECH-RA--1994-1996
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 29040477
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BENCHMARKS; CORRELATION FUNCTIONS; MULTIPARTICLE SPECTROMETERS; POSITION SENSITIVE DETECTORS; SI SEMICONDUCTOR DETECTORS; SPATIAL RESOLUTION
- Descriptors DEC
- FUNCTIONS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RESOLUTION; SEMICONDUCTOR DETECTORS; SPECTROMETERS
Optional Information
- Notes
- 2 refs. Imprint:Rapport d'activite, 1994 - 1996