Published January 2001 | Version v1
Report

Optimising the design of gas microstrip detectors for soft x-ray detection

Description

This report describes development work in which systematic changes in the electrode pattern of a Gas Microstrip Detector are explored in the search for higher avalanche gains and enhanced stability. It is found that the width of the cathode structure is the main determinant of the detector stability. With the correct cathode width, gas gains of >50 000 are comfortably attainable with low detector noise so that x-rays can potentially be detected down to the limit of a single x-ray-produced photoelectron. (author)

Availability note (English)

Available from British Library Document Supply Centre- DSC:8715.1804(2001-008)

Additional details

Publishing Information

Imprint Pagination
12 p.
Journal Issue
no.RAL-TR-2001-008
Series
Technical report
ISSN
1358-6254
Report number
CLRC-RAL-TR--2001-008

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
32010533
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Non-conventional Literature
Descriptors DEI
CATHODES; PHOTOELECTRON COUNTING; STABILITY; TOWNSEND DISCHARGE; X-RAY DETECTION
Descriptors DEC
COUNTING TECHNIQUES; DETECTION; ELECTRIC DISCHARGES; ELECTRODES; RADIATION DETECTION