Published January 2001
| Version v1
Report
Optimising the design of gas microstrip detectors for soft x-ray detection
Description
This report describes development work in which systematic changes in the electrode pattern of a Gas Microstrip Detector are explored in the search for higher avalanche gains and enhanced stability. It is found that the width of the cathode structure is the main determinant of the detector stability. With the correct cathode width, gas gains of >50 000 are comfortably attainable with low detector noise so that x-rays can potentially be detected down to the limit of a single x-ray-produced photoelectron. (author)
Availability note (English)
Available from British Library Document Supply Centre- DSC:8715.1804(2001-008)Additional details
Publishing Information
- Imprint Pagination
- 12 p.
- Journal Issue
- no.RAL-TR-2001-008
- Series
- Technical report
- ISSN
- 1358-6254
- Report number
- CLRC-RAL-TR--2001-008
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 32010533
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Non-conventional Literature
- Descriptors DEI
- CATHODES; PHOTOELECTRON COUNTING; STABILITY; TOWNSEND DISCHARGE; X-RAY DETECTION
- Descriptors DEC
- COUNTING TECHNIQUES; DETECTION; ELECTRIC DISCHARGES; ELECTRODES; RADIATION DETECTION