Published 2004 | Version v1
Miscellaneous

Monte Carlo simulation of the dose distribution around the 125I model 6711 seed as function of radius of the silver cylinder using the Penelope code

  • 1. Instituto Nacional de Cancerologia, Bogota (Colombia)
  • 2. Universidad de Cordoba, Monteria (Colombia)
  • 3. Universidad Nacional de Colombia, Bogota (Colombia)
  • 4. Universite de la Mediterranee, Marseille (France)

Description

The Monte Carlo method is applied to find the dose rates distribution in tissue around 125 I seeds model 6711 as a function of the silver cylinder radius, Rsc (0.017, 0.021, 0.025, 0.029 and 0.033) cm are used as radius values. It is found here that the dose rate at any point within the tissue decreases as Rsc increases. The relative difference of dose rate that produced by the standard Rsc seed, is less than 5%, for seeds with Rsc between 0.017 and 0.033 cm. (author)

Part of:
Proceedings of the 9. Brazilian congress on medical physics; 3. Iberian Latin American and Caribbean congress on medical physics

Additional details

Additional titles

Original title (Spanish)
Simulacion por el Metodo de Monte Carlo de la distribucion de dosis alrededor de la semilla de

Identifiers

Publishing Information

Imprint Title
Proceedings of the 9. Brazilian congress on medical physics; 3. Iberian Latin American and Caribbean congress on medical physics
Imprint Pagination
[vp.]
Journal Page Range
1 p.

Conference

Title
9. Brazilian congress on medical physics; 3. Iberian Latin American and Caribbean congress on medical physics
Original Conference Title
9. Congresso brasileiro de fisica medica
Dates
26-29 Sep 2004
Place
Rio de Janeiro, RJ (Brazil)

Optional Information

Notes
14 refs., 3 figs., 1 tab. Imprint:Anais do 9. Congresso brasileiro de fisica medica; 3. Congresso ibero latino americano e do Caribe de fisica medica