Noise analysis for calorimetric low-temperature detectors for heavy ions
Description
The energy resolution of calorimetric low-temperature detectors for heavy ions has been analyzed. It is shown that the contribution of base line noise is small. The energy resolution is determined by intrinsic fluctuations of the detector signal. An incomplete energy thermalization during the stopping process of the heavy ion, the dependence of signal shape on impact position and fluctuations of the Al-TES thermometer response are considered as main sources of detector line broadening. Test measurements with 5 MeV α-particles are presented
Additional details
Identifiers
- DOI
- 10.1016/j.nima.2003.11.227;
- PII
- S0168900203031292;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 520
- Journal Issue
- 1-3
- Journal Page Range
- p. 84-86
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 10. international workshop on low temperature detectors
- Dates
- 7-11 Jul 2003
- Place
- Genoa (Italy)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 35080976
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ENERGY RESOLUTION; FLUCTUATIONS; HEAVY IONS; LINE BROADENING; MEV RANGE 01-10; PHONONS; RADIATION DETECTION; RADIATION DETECTORS; TEMPERATURE RANGE 0065-0273 K; THERMALIZATION; THERMOMETERS; VENTILATION BARRIERS
- Descriptors DEC
- CHARGED PARTICLES; DETECTION; ENERGY RANGE; ENGINEERED SAFETY SYSTEMS; IONS; MEASURING INSTRUMENTS; MEV RANGE; QUASI PARTICLES; RESOLUTION; SLOWING-DOWN; TEMPERATURE RANGE; VARIATIONS
Optional Information
- Copyright
- Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.