Published March 11, 2013 | Version v1
Journal article

Position mapping and a uniformity correction method for small-animal SPECT based on connected regional recognition

  • 1. Department of Biomedicine and Engineering, College of Engineering, Peking University, No. 5 Yiheyuan Road, Beijing 100871 (China)
  • 2. Department of Control Technology and Instrument, College of Quality and Technical Supervision, Hebei University, No.180 Wusi East Road, Baoding 071000 (China)

Description

We describe a novel position mapping and the uniformity correction method to improve the performance of small animal single-photon emission computed tomography (SPECT) imaging. The SPECT system consists of a cerium doped lutetium−yttrium oxyorthosilicate (LYSO) scintillation crystal (22×22 pixel array, 2 mm×2 mm×3 mm pixel size), a position sensitive photomultiplier tube (H8500c, Hamamatsu Photonics Co., Ltd., Shizuoka Prefecture, Japan), and a parallel-hole collimator (Nuclear Fields Pty. Ltd., St. Marys, Australia). The position mapping method was based on a connected regional recognition algorithm. We present the algorithm and step-by-step details of image boundary detection, dynamic binarization, connected regional recognition, center-of-gravity computing, and look-up table establishment. The position mapping and uniformity correction tables were generated and applied to the SPECT projection data. The corrected projection images demonstrated that this correction method improved the uniformity of the raw projection image by ∼16%. The preliminary SPECT reconstruction results (using algebraic reconstruction technology) are also presented. A comparison between the reconstructed images before and after correction further confirms the performance of this correction method

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2012.11.008

Additional details

Identifiers

DOI
10.1016/j.nima.2012.11.008;
PII
S0168-9002(12)01300-9;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
704
Journal Issue
Complete
Journal Page Range
p. 1-6
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2012 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.