Published April 2013 | Version v1
Journal article

Dose impact of a carbon fiber couch for stereotactic body radiation therapy of lung tumors

  • 1. Kumamoto Radiosurgery Clinic, Kumamoto, Kumamoto (Japan)
  • 2. Kumamoto Univ., Faculty of Life Sciences, Kumamoto, Kumamoto (Japan)
  • 3. Kumamoto Univ. Hospital, Kumamoto, Kumamoto (Japan)
  • 4. Kumamoto Univ., Graduate School of Health Sciences, Kumamoto, Kumamoto (Japan)

Description

The aim of this study was to measure the dose attenuation caused by a carbon fiber radiation therapy table (Imaging Couch Top; ICT, BrainLab) and to evaluate the dosimetric impact of ICT during stereotactic body radiation therapy (SBRT) in lung tumors. The dose attenuation of ICT was measured using an ionization chamber and modeled by means of a treatment planning system (TPS). SBRT was planned with and without ICT in a lung tumor phantom and ten cases of clinical lung tumors. The results were analyzed from isocenter doses and a dose-volume histogram (DVH): D95, Dmean, V20, V5, homogeneity index (HI), and conformity index (CI). The dose attenuation of the ICT modeled with TPS agreed to within ±1% of the actually measured values. The isocenter doses, D95 and Dmean with and without ICT showed differences of 4.1-5% for posterior single field and three fields in the phantom study, and differences of 0.6-2.4% for five fields and rotation in the phantom study and six fields in ten clinical cases. The dose impact of ICT was not significant for five or more fields in SBRT. It is thus possible to reduce the dose effect of ICT by modifying the beam angle and beam weight in the treatment plan. (author)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi
Journal Volume
69
Journal Issue
4
Journal Page Range
p. 400-406
ISSN
0369-4305