Published February 21, 2005 | Version v1
Journal article

Radiation sensitivity of GSO and LSO scintillation detectors

  • 1. Institute of Technological Investigations, 25163 Strancice, Spojovaci 377, Prague (Czech Republic)
  • 2. SECUREX Ltd., Hi-Tech and Engineering Co., 25163 Strancice, Prague (Czech Republic)

Description

Radiation resistance of 4x4x30 mm3 GSO and LSO imaging scintillation detectors has been studied for low-energy gamma-ray doses of 104 Gy (106 rad) and 105 Gy (107 rad). Radiation hardness was determined by the measurement of optical transmission through GSO and LSO scintillation crystals before and after irradiations with 60Co gamma-rays. The results have been analysed in terms of the radiation-induced absorption coefficients and compared with radiation sensitivity measurements of small BGO scintillation crystals. The recovery time of irradiated small GSO and LSO crystals has also been determined

Additional details

Identifiers

DOI
10.1016/j.nima.2004.10.014;
PII
S0168-9002(04)02271-5;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
539
Journal Issue
1-2
Journal Page Range
p. 132-136
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2004 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.