Published February 1978
| Version v1
Journal article
Internal quality control of radioimmunoassays
Description
A modified cumulative sum technique has been applied to radioimmunoassay quality control data. The method is approximately 50% more efficient in detecting systematic changes in the mean and variance of quality control values for plasma samples than more widely used conventional methods. The salient features of the technique have been restricted to changes in the mean quality control value of a plasma pool, but potential applications to changes in variance and as a diagnostic aid to problems in radioimmunoassay have been evaluated. The method is independent of computing facilities and statistical expertise since all computations have been presented in the form of a nomogram and thus can be used by technicians at the bench. (author)
Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Endocrinology
- Journal Volume
- 76
- Journal Issue
- 2
- Series
- J. Endocrinol.
- Journal Page Range
- 203-210
- ISSN
- 1479-6805
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 9395236
- Subject category
- S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- BLOOD PLASMA; ERRORS; FLUCTUATIONS; NOMOGRAMS; PERFORMANCE TESTING; QUALITY CONTROL; RADIOIMMUNOASSAY
- Descriptors DEC
- BIOLOGICAL MATERIALS; BLOOD; BODY FLUIDS; CONTROL; DIAGRAMS; INFORMATION; ISOTOPE APPLICATIONS; TESTING; TRACER TECHNIQUES; VARIATIONS
Optional Information
- Notes
- Updated automatically by Metadata and Full-Text Enrichment Agent