Radiation exposure to angiographers under different fluoroscopic imaging conditions

  • 1. Thomas Jefferson University, Philadelphia, PA (USA)

Description

Radiation levels near an imaging chain commonly used in angiography were measured with both a 100- and a 200-mm-thick scatter phantom. The scatter was measured in lines parallel in space to the central ray of the x-ray beam, at lateral distances of 300, 500, and 800 mm. The effects of fluoroscopic kilovoltage and image intensifier magnification mode were also measured. The results indicate that the highest scattered radiation levels occur near the surface of the patient where the x-ray beam enters. Exposure rates were measured in both anteroposterior (AP) and posteroanterior (PA) geometries on a U-arm system. In PA geometry, the highest radiation levels occur below the angiographer's waist, an area well protected by the lead apron. The AP geometry increases the exposure rate to the neck, head, and upper extremities, areas where apron shielding is less effective

Additional details

Publishing Information

Journal Title
Radiology
Journal Volume
180
Journal Issue
3
Series
Radiology.
Journal Page Range
861-865
ISSN
0033-8419
CODEN
RADLA